尝试输入关键词、订货号/产品型号或序列号(“CM442”或“技术资料”)进行查找
至少输入2个字符开始搜索。
单通道H2O或CO2气体分析仪SS2000e的产品图(正视图)

SS2000E

SS2000E
Please Wait (spinning wheel)
价格核算中…
暂时无价格信息
Price on request
正在激活价格
价格:提供信息
筛选测量/应用任务
测量/应用任务

配置产品,下载CAD设计图

  • 测量原理

    TDLAS

  • Product Headline

    Single channel gas analyzer for moisture (H2O) in the natural gas industry. The sensor measures gas using tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. Includes a heated enclosure for the sample conditioning system.

  • Channels

    1

  • Analyte and Measurement ranges

    H2O (Moisture):
    0 to 100 ppmv (minimum)
    0 to 5000 ppmv (maximum)

  • Measured Variables

    Concentration
    Moisture Dew Point
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 700 to 1400 mbara or 700 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    100 to 240 VAC, 50 to 60 HZ
    or
    18 to 24 VDC - optional
    1 Amp maximum at 120 VAC, 1.6 Amps at 24 VDC (unheated)
    2 Amps maximum at 120 VAC (heated)

  • Communication

    Analog: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C - standard, RS485 and Ethernet - optional
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 2, General Fault and Concentration/Assignable Alarm

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 4X 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    Type 4X

  • 测量原理

    TDLAS

  • Product Headline

    Single channel gas analyzer for carbon dioxide (CO2) in the natural gas industry. The sensor measures gas using tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. Includes a heated enclosure for the sample conditioning system.

  • Channels

    1

  • Analyte and Measurement ranges

    CO2 (Carbon Dioxide):
    0 to 5% volume (minimum)
    0 to 50% volume (maximum)

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 700 to 1400 mbara or 700 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    100 to 240 VAC, 50 to 60 HZ
    or
    18 to 24 VDC - optional
    1 Amp maximum at 120 VAC , 1.6 Amps at 24 VDC (unheated)
    2 Amps maximum at 120 VAC (heated)

  • Communication

    Analog: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C - standard, RS485 and Ethernet - optional
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 2, General Fault and Concentration/Assignable Alarm

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 4X 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    Type 4X